3 exchanging the electronics module, 4 software update – VEGA VEGAPULS 67 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART two-wire User Manual
Page 48
8
.3 Exchanging the electronics module
I
f the electronics module is defective, it can be replaced by the user.
I
n Ex applications only one instrument and one electronics module with
respective Ex approval may be used.
I
f there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.
T
he new electronics module must be loaded with the settings of the
sensor. These are the options:
l
A
t the factory by VEGA
l
O
r on site by the user
I
n both cases, the sensor serial number is necessary. The serial
numbers are stated on the type label of the instrument, inside the
housing or on the delivery note.
Information:
W
hen loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Oscillator").
T
he electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
8
.4 Software update
T
he software version of VEGAPULS 67 can be determined as follows:
l
on the type label of the electronics
l
via the indicating and adjustment module
l
via PACTware
Y
ou can view all software histories on our website www.vega.com.
M
ake use of this advantage and get registered for update information
via e-mail.
T
he following components are required to update the sensor software:
l
S
ensor
l
V
oltage supply
l
VEGACONNECT
l
PC
with PACTware
l
C
urrent sensor software as file
A
t "www.vega.com/downloads" go to "Software". Select under "plics
instruments and sensors" the suitable instrument series. Load the zip
fi
le via the right mouse key with "Save target as" e.g. on the desktop of
your PC. Extract all files available in the zip file, e.g. to the desktop.
Sensor serial number
Assignment
Load sensor software to
PC
48
VEGAPULS
67 • 4 … 20
mA/HART two-wire
8 M
aintenance and fault rectification
32938
-EN
-100514