beautypg.com

VEGA VEGAFLEX 66 (-200…+400°C) 4 … 20 mA_HART two-wire User Manual

Page 60

background image

-

EE

x-ia instrument

20 … 30 V DC

-

EE

x-d-ia instrument

20 … 36 V DC

P

ermissible residual ripple

-

<

100 H

z

U

ss

<

1 V

-

100 H

z … 10 kHz

U

ss

<

10

mV

L

oad

see diagram

1000

750

500

250

14

18

16

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 35: Voltage diagram

1

HART load

2

Voltage limit EEx-ia instrument

3

Voltage limit non-Ex/Ex-d instrument

4

Operating voltage

Electrical protective measures

P

rotection, depending on housing version

-

P

lastic housing

IP

66/IP 67

-

A

luminium housing, stainless steel

housing - investment casting, stainless
steel housing - electro-polished

IP

66/IP 68 (0.2

bar)

4)

-

A

luminium and stainless housing, in-

vestment casting (optionally available)

IP

66/IP 68 (1

bar)

O

vervoltage category

III

P

rotection class

II

Functional safety (SIL)

F

unctional safety is already activated on instruments with SIL qualification ex factory. On

instruments without SIL qualification ex factory, the functional safety must be activated by the user
via the indicating and adjustment module or via PACTware for applications according to SIL.

F

unctional safety according to IEC 61508-4

-

S

ingle channel architecture (1oo1D)

up to SIL2

4)

A

suitable cable is the prerequisite for maintaining the protection rating.

60

VEGAFLEX

66 • 4 … 20

mA/HART two-wire

10 S

upplement

34173

-EN

-100427