beautypg.com

2 mounting instructions, 3 mounting steps – VEGA VEGABAR 63 4 … 20 mA_HART User Manual

Page 14

background image

1

2

Fig. 4: Temperature ranges
1

Process temperature

2

Ambient temperature

Danger:
I

nstruments in the version "Oil and grease free for oxygen" should be

unpacked just before mounting. After the protective cover of the
process fitting has been removed, the label "O₂" on the process fitting
is visible. Contamination by oil, grease and dirt should be avoided.
D

anger of explosion!

4.2 Mounting instructions

VEGABAR

63

functions in any installation position. It is mounted

according to the same directives as a manometer (DIN EN 839-2).

Information:
W

e recommend using lock fittings, measuring instrument holders and

siphons from our line of accessories.

4.3 Mounting steps

F

or mounting VEGABAR 63, a welded socket is required. You can find

these components in the supplementary instructions manual "Welded
socket and seals
".

U

se the seal corresponding to the instrument:

l

P

rocess fitting GV, GF, GC: seal in front of the thread

-

or -

S

eal the thread with teflon, hemp or a similar resistant seal material:

l

P

rocess fitting GN

Oxygen applica-
tions

Mounting position

Welding the socket

Sealing/Screwing in

14

VEGABAR

63 • 4 … 20

mA/HART

4 M

ounting

32462

-EN

-081209