Fig. 21: Exchanging the cable or rod

Information:
Please keep the stated torque so that the max. tensile strength
remains.

The cable and rod of the probe can be shortened by any amount.

1

Cable: loosen the three pins on the gravity weight (hexagon 3)

2

Cable: remove the pins

3

Cable: Pull the cable out of the gravity weight

4

Shorten the cable/rod with a cut-off wheel or metal saw at the
lower end. Make sure the length is correct.

5

Cable: cable protrudes approx. 40 mm (1.575 in) into the gravity
weight.

6

Cable: Fasten the cable with three pins, torque 7 Nm (5.16 lbf ft)

7

Enter new probe length and then carry out a fresh adjustment (see
"S

etup procedure, Carrying out min. adjustment - Carrying out

max. adjustment").

Shorten cable/rod