beautypg.com

VEGA VEGACAP 62 - Relay (DPDT) User Manual

Page 23

background image

23

7 Maintenance and fault rectification

VEGACAP 62 • - Relay (DPDT)

30004-EN-130916

Error

Cause

Rectification

– The instrument

signals covered

without cover-

ing with the

medium

– The instrument

signals covered

with cover-

ing with the

medium

Wrong mode se-

lected on the

signal condition-

ing instrument

Set the correct mode on the mode

switch of the signal conditioning in-

strument (A: overflow protection, B:

dry run protection). Wiring should be

carried out according to the quiescent

current principle.

Operating voltage

too low

Check operating voltage

Shortcircuit in the

probe, e.g. be-

cause of moisture

in the housing

Remove the electronics module.

Check the resistance between the

marked plug connections. See the fol-

lowing instructions.

Electronics de-

fective

Push the mode switch (A/B) on the

signal conditioning instrument. If the

signal conditioning instrument then

changes the mode, the probe may

be mechanically damaged. Should

the switching function in the correct

mode still be faulty, return the probe

for repair.
Check if there is buildup on the probe,

and if so, remove it.

Unfavourable in-

stallation location

Check if the probe is covered by build-

up on the socket.
Mount the instrument at a location

in the vessel where e.g. no mounds

can form.

Signal lamp flash-

es red

Electronics mod-

ule has detected a

failure

Exchange the instrument or send it in

for repair

Remove the electronics module. Check the resistance between the

two plug connections.
There must no longer be a connection (high impedance). If there is

still a connection - exchange the instrument or return it for repair

Checking the switching

signal

Check the resistance in

the probe