Reference conditions to measuring accuracy (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
-
Temperature
+
18 … +30 °C (+64 … +86 °F)
-
Relative humidity
45 … 75 %
-
Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Other reference conditions
-
Reflector
ideal reflector, e.g. metal plate 2 x 2 m
-
False reflections
Biggest false echo, 20 dB smaller than the useful
echo
Characteristics and performance data
Frequency
C-band (6 GHz technology)
Step response or adjustment time
8)
>
1 s (dependent on the parameter setting)
Max. level change
adjustable up to 1 m/min. (dependent on the
parameter adjustment)
Max. emitted HF power of the antenna system
-
Pulse peak power approx.
0.1 mW
-
Pulse duration
<
2 ns
-
SAR value
9)
0.471 mW/kg
Measuring accuracy
Resolution, general
max. 1 mm (0.039 in)
Deviation
10)
see diagrams
0,5 m
(1.6 ft)
5,85 m
(19.19 ft)
10 mm (0.394 in)
5 mm (0.197 in)
- 5 mm (- 0.197 in)
- 10 mm (- 0.394 in)
0
Fig. 24: Deviation VEGAPULS 66 standpipe
8)
Time to output the correct level (with max. 10 % deviation) after a sudden
level change.
9)
at the aperture of a horn antenna with ø 150 mm
10)
Incl. non-linearity, hysteresis and non-repeatability.
VEGAPULS
66 standpipe version • Profibus PA
45
10 Supplement
28743
-EN
-090305