beautypg.com

VEGA VEGABAR 65 4 … 20 mA_HART User Manual

Page 61

background image

C

onnection cable between IP 68 instrument and external housing:

-

C

onfiguration

four wires, one suspension cable, one breather
capillary, screen braiding, metal foil, mantle

-

W

ire cross-section

0.5

mm² (AWG 20)

-

W

ire resistance

<

0.036 Ω/

m (0.011 Ω/ft)

-

S

tandard length

5

m (16.40 ft)

-

M

ax. length

180

m (590.5 ft)

-

M

in. bending radius at 25 °C/77 °F

25

mm (0.985 in)

-

D

iameter approx.

8

mm (0.315 in)

-

C

olour - standard PE

B

lack

-

C

olour - standard PUR

B

lue

-

C

olour - Ex-version

B

lue

Indicating and adjustment module

V

oltage supply and data transmission

through the sensor

I

ndication

LC

display in dot matrix

A

djustment elements

4

keys

P

rotection class

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

M

aterials

-

H

ousing

ABS

-

I

nspection window

P

olyester foil

Power supply

O

perating voltage

-

N

on-Ex instrument

12 … 36 V DC

-

EE

x-ia instrument

12 … 30 V DC

-

E

xd instrument

18 … 36 V DC

O

perating voltage with lighted indicating and adjustment module

-

N

on-Ex instrument

20 … 36 V DC

-

EE

x-ia instrument

20 … 30 V DC

-

EE

x-d-ia instrument

20 … 36 V DC

P

ermissible residual ripple

-

<

100 H

z

U

ss

<

1 V

-

100 H

z … 10 kHz

U

ss

<

10

mV

L

oad

see diagram

VEGABAR

65 • 4 … 20

mA/HART

61

10 S

upplement

27531

-

EN

-081209