Dynamic behaviour output

Run-up time approx.

10 s

90 %

100 %

10 %

t

t

T

A

t

t

S

2

1

Fig. 28: Sudden change of the process variable, dead time t

T

, rise time t

A

and step response time t

S

1

Process variable

2

Output signal

Dead time

150 ms

Rise time

100 ms (10 … 90 %)

Step response time

250 ms (ti: 0 s, 10 … 90 %)

To this amounts the reaction time of the isolatng system. This time varies from values < 1 s with
compact isolating diaphragms up to several seconds with capillary systems.

Example: Flange isolating diaphragm DN 80, filling silicone oil KN 2.2, capillary length 10 m,
measuring range 1 bar

Process temperature

Reaction time

40 °C

approx. 2 s

20 °C

approx. 3 s

-20 °C

approx. 11 s

Damping (63 % of the input variable)

0 … 999 s, adjustable

Input variable

Adjustment

Adjustment range of the min./max. adjustment relating to the nominal measuring range:

-

percentage value

-10 … 110 %

-

pressure value

-20 … 120 %