VEGA VEGABAR 61 Profibus PA User Manual
Page 36
6
Enter the requested density value with [->] and [+], confirm with
[OK]
and move to position correction with [->].
The adjustment unit is hence switched over from bar to m.
Proceed as follows to select the temperature unit:
5)
à Activate the selection with [OK] and select "Temperature unit
with [->].
à Activate the selection with [OK] and select the requested unit with
[->]
(e.g. °F).
à Confirm with [OK].
The temperature unit is hence switched over from °C to °F.
Proceed as follows:
1
Activate in the menu item "Position correction" the selection with
[OK]
.
Position correction
Offset
P
=
+
0000 mbar
53 mbar
2
Select with [->], e.g. to accept actual measured value.
Position correction
Accept current measured
value?
▶
Accept
Edit
3
Confirm with [OK] and move to min.(zero) adjustment with [->].
Proceed as follows:
1
Edit the % value in the menu item "Min. adjustment" with [OK].
Min. adjustment
+000
.0 %
=
+0000
.0 mbar
0000.0 mbar
2
Set the requested percentage value with [+] and [->].
3
Edit the requested mbar value with [OK].
4
Set the requested mbar value with [+] and [->].
5
Confirm with [+] and move to max. adjustment with [->].
The min. adjustment is finished.
5)
Selection options: °C, °F.
Carry out position cor-
rection
Carrying out min. ad-
justment
36
VEGABAR
61 • Profibus PA
6 Set up with the indicating and adjustment module PLICSCOM
28251
-EN
-081209