VEGA VEGAPULS 66 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA_HART four-wire User Manual
Page 48
Reference conditions to measuring accuracy (similar to DIN EN 60770-1)
R
eference conditions according to DIN EN 61298-1
-
T
emperature
+
18 … +30 °C (+64 … +86 °F)
-
R
elative humidity
45 … 75 %
-
A
ir pressure
860 … 1060
mbar/86 … 106 kPa (12.5 … 15.4 psig)
O
ther reference conditions
-
R
eflector
ideal reflector, e.g. metal plate 2 x 2 m
-
F
alse reflections
B
iggest false echo, 20 dB smaller than the useful
echo
Characteristics and performance data
F
requency
C-
band (6 GHz technology)
B
eam angle 3 dB, depending on the antenna system
-
without horn antenna
38°
-
ø
75
mm (2.953 in)
38°
-
ø
96
mm (3.78 in)
30°
-
ø
146
mm (5.748 in)
20°
-
ø
196
mm (7.717 in)
17°
-
ø
242
mm (9.528 in)
14°
S
tep response or adjustment time
4)
>
1
s (dependent on the parameter setting)
M
ax. level change
adjustable up to 1 m/min. (dependent on the
parameter adjustment)
M
ax. emitted HF power of the antenna system
-
P
ulse peak power approx.
0.1
mW
-
P
ulse duration
<
2
ns
-
SAR
value
5)
0.471
mW/kg
Measuring accuracy
R
esolution, general
max. 1 mm (0.039 in)
D
eviation
6)
see diagrams
4)
T
ime to output the correct level (with max. 10 % deviation) after a sudden
level change.
5)
at the aperture of a horn antenna with ø 150 mm
6)
I
ncl. non-linearity, hysteresis and non-repeatability.
48
VEGAPULS
66 • 4 … 20
mA/HART four-wire
10 S
upplement
28443
-EN
-090305