1.5 Technical data

General

Series

instrument for front panel or wall mounting or
mounting on rail 35 x 7,5

Dimensions

W = 96 mm, H = 104 mm, D = 90 mm

Weight

approx. 400 g

Housing material

plastic ABS/POM

Cross section area of conductor

max. 2,5 mm

2

Ambient conditions

Permissible ambient temperature

-20°C … +60°C

Storage and transport temperature

-40°C … +85°C

Temperature error

0,01 %/K

Electrical protective measures

Protection class

II

Overvoltage category

II

Protection
- wall or carrier rail mounting

IP 20

- front panel mounting

IP 40

Voltage supply

Supply voltage

20 … 250 V AC/DC

Power consumption

4 W, 12,5 VA

Sensor input (floating)

Transmission

analogue

Connection line to the sensor

2-wire

I

active

- sensor supply

25 … 15 V (range 4 ... 20 mA)

- shortcircuit limitation

approx. 30 mA

I

passive

- inner resistance

< 250

Ω (range 4 ... 20 mA)

U
- inner resistance

> 100 k

Ω (range 0 ... 10 V)

Current output 0/4 … 20 mA (signal circuit, floating)

Resolution

0,1 % (range 0/4 ... 20 mA)

max. load

500

Ω

Fault signal

approx. 22 mA

Linearity error

0,1 %

Relay output

Number of relays

4 (2 modules with 2 relays each)

Turn-on voltage

min. 10 mV
max. 250 V AC, 250 V DC

Switching current

min. 10 μA
max. 3 A AC, 1 A DC

Breaking capacity

max. 54 W DC, 500 VA AC

Indication of switching condition

LED lights - relay energized
LED off - relay deenergized

Digital indication

Zero point (4 mA or 0 V)

-9999 … 9999

Final point (20 mA or 10 V)

-9999 … 9999

Decimal point