6.1 Sensor Calibration
The input sensitivity can be calibrated using the calibration function and a sensor calibrator. Click the
button on the Setup interface to launch the Calibration interface.
Step 1
Adjust the calibration parameters
The defaults values are:
• Averaging time:
5 s
• Frequency:
1 kHz
• Calibrator Level:
94 dB
Step 2
Install the calibrator device on the microphone
Step 3
Click START
After the average time is elapsed, the sensitivity value will update.
Step 4
Click OK to accept the sensitivity value