20
Shape Test Numerical Results at 1 kHz
The following table presents the numerical results of the shape test at 1 kHz:
Frequency (Hz)
Low limit (dB)
Measurement (dB)
High limit (dB)
185.5
-inf
-96.0
-75.0
327.5
-inf
-85.1
-62.0
531.4
-inf
-61.1
-42.5
772.6
-inf
-28.2
-18.0
891.3
-4.5
-3.0
-2.3
919.6
-1.1
-0.3
0.15
947.0
-0.4
0.0
0.15
974.0
-0.2
0.0
0.15
1000.0
-0.15
0.0
0.15
1026.7
-0.2
0.0
0.15
1055.8
-0.4
0.0
0.15
1087.5
-1.1
-0.3
0.15
1122.0
-4.5
-3.0
-2.3
1294.4
-inf
-28.4
-18.0
1881.7
-inf
-61.8
-42.5
3053.7
-inf
-86.9
-62.0
5392.0
-inf
-96.7
-75.0