6. MAINTENANCE

6.1 Troubleshooting

The following table lists common problems and possible causes and remedies. If the
problem persists, contact your authorized dealer.

TABLE 6-1, TROUBLESHOOTING.

Symptom Possible

Cause

Remedy

Cannot turn on

• No power to scale
• Empty or incorrectly installed

batteries

• Verify AC adapter connections and voltage.
• Replace batteries, check polarity.

Poor accuracy

• Improper calibration
• Unstable environment
• Transportation Lock is set
• Debris touching the pan
• Draft shield is touching the pan

or sample.

• Perform calibration.
• Move scale to suitable location.
• Slide the pin to the unlocked position.
• Clean any debris next to or under the pan.
• Check that pan is fully seated, check sample height.

Cannot calibrate

• Unstable environment
• Incorrect calibration mass
• Transportation Lock is set
• Calibration Lock switch is set

• Move the scale to a suitable location.
• Use correct calibration masses.
• Slide the pin to the unlocked position.
• Slide the switch to the unlocked position.

Err 1

• Invalid Checksum data.

• Cycle scale off/on.
• Contact your authorized dealer.

Err 2

• Over-load or Under-load

condition.

• Transportation Lock is set
• Improper calibration

• Verify pan is installed and/or excess load is

removed from pan.

• Slide the pin to the unlocked position.
• Perform calibration.

Err 4

• Invalid calibration mass.

• Verify correct calibration mass is being used.

Err 5

• The titration test was performed

incorrectly or the PPM result
was >99999.99

• Repeat the titration test and perform the correct

steps in the proper order.

• Concentrations >99999.99 cannot be tested.

Err 8

• RS232 buffer is full.

• Cycle scale Off/On.
• Check RS232 settings in scale and computer when

RS232 option is installed.

Err 9

• Internal data error.

• Contact your authorized dealer.

(Flashing)

• Battery power is low.

• Replace batteries soon.

bAtt LowW

• Battery power is too low for

proper operation.

• Replace batteries.

6.2 Service Information

If the troubleshooting section does not resolve or describe your problem, contact your
authorized dealer.