Ohaus GraviTek Titration Scale User Manual
Page 12

GraviTek™ Titration Scale
EN-10
flash.
4.2.2 Type/Unit Menu
[t.Y.P.E]
The Type/Unit menu contains sub-menus to separately enable or disable each
available Type/Unit.
Enable Type/Unit [Unit
xx
]:
Settings are On OFF. See Table 4-1 for the
symbols
associated with each Type/Unit.
4.2.3 Print Menu
The print menu is only available only when the USB option is installed. The Print
menu contains the following sub-menus:
Stable Only Print [StAbLE]: Stable Only Print settings: Off On
Setting to [On] will allow only stable weight data to be
printed. Stable weight data is shown by the state of the
Stability indicator [
*
].
Auto Print [APrint]:
Auto Print settings: Off Cont on.StbL on.tESt 5
SEC
15 SEC 30 SEC 60 SEC
.
All Auto Print settings (except on.tESt) affect printing
only while in the standard weighing mode.
[Cont] Continuous Print sends weight data as fast as
possible.
[on.StbL] On Stability Print sends weight data each time
a new stable reading is obtained. The weight change must
be greater then 5d and greater then 5d from zero.
[on.tESt] On Test Print operates only in the Titrate
Mode. A single line print out is sent at the completion of a
PPM calculation. Three values are sent in a single line
format.
[
XX
SEC
] Interval Print sends weight data at the selected
XX
second interval with a relative time stamp appended to
each print out.
Note: In all Auto Print modes other then [on.tESt]
pressing the Print button will cause a temporary stop to