beautypg.com

Ohaus GraviTek Titration Scale User Manual

Page 11

background image

GraviTek™ Titration Scale

EN-9

next setting press No. To accept the displayed setting press Yes.


Menu Exit:

The last item in each sub-menu list is [EnD]. Press Yes to return
to the main menu. The last item in the main menu is also
[E.n.D]. Press Yes to exit the main menu and return to the
previous weighing mode.


4.2.1 Setup Menu

[S.E.t.U.P.]

The Setup menu contains the following sub-menus:

Auto-Off [A-OFF]: Auto

Off

settings:

Off On

Setting to [On] will automatically turn the scale off after no
activity is detected for 6 minutes.


Backlight [LIGHt]: Backlight

settings:

Off On Auto

Setting to [Auto] will allow the Backlight to turn on with
any activity. Then after 5 seconds of no activity the
Backlight will automatically turn off.


Span Calibration [SPAN]: Press Yes to begin Span calibration. Span calibration uses

two weight readings: Zero and a value at or near the full
capacity of the scale. Span calibration should be performed
if the scale is repositioned or if the room temperature
changes significantly.


Linearity Calibration [Lin]: Press Yes to begin Linearity calibration. Linearity

calibration uses three weight readings: Zero, a mid-range
value and full capacity. Generally Linearity calibration is
not required unless testing shows the linearity error exceeds
the tolerance in the specification table. Accuracy of the
calibration weights is important to maintain weighing
integrity.


Display Memory [dSPY.mME]: Press Yes to begin the display of stored test memory.

Press No at any time to exit the display memory function.
Pressing Yes repeatedly will advance to the next test that is
stored in memory.


Clear Memory [CLr.mME]:

Press Yes to erase all stored tests in the memory.

The display will flash one time when memory is erased. If
there is no test memory to be erased the display will not