beautypg.com

En-9 pioneer, 2 specifications, 3 communication – Ohaus PIONEER ANALYTICAL AND PRECISION BALANCES Manual fr User Manual

Page 11

background image

EN-9

Pioneer

6.2 Specifications

*C= Internal calibration

6.3 Communication

6.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed

weight. The balance will return “ES” for invalid commands.

Command Function

IP

Immediate Print of displayed weight.

P

Print displayed weight (uses Stable ON/OFF menu settings).

CP

Continuous Print.

xP

Interval Print x = Print Interval (1-3600 sec)

T

Same as pressing Zero Key.

ON

Turns balance ON.

OFF

Turns balance OFF.

PSN

Show Serial Number.

PV

Version: Print product name, software revision and LFT ON (if LFT is set ON).

PU

Print current mode/unit

x#

Set PC ref wt (x) in grams

P#

Print PC ref wt

x%

Set % ref wt (x) in grams

P%

Print % ref wt

6.3.2 RS232 (DB9) Pin Connections

Pin 2: Balance transmit line (TxD)

Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)

Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)

MODEL

PA64

*

PA114

*

PA214

*

PA213

*

PA413

*

PA512

*

PA2102

*

PA4102

*

PA4101

*

Capacity (g)

65

110

210

210

410

510

2100

4100

4100

Readability (g)

0.0001

0.001

0.01

0.1

Repeatability (sd)

0.0001

0.001

0.01

0.1

Linearity (g)

± 0.0002

± 0.002

± 0.02

± 0.1

Tare Range

To capacity by subtraction

Stabilization

3 seconds

Span cal mass (g)

50 or

60

50 or

100

100 or

200

100 or

200

200 or

400

200 or

500

1000 or

2000

2000 or

4000

2000 or

4000

Lin. cal masses (g)

25,

50

50,

100

100,

200

100,

200

200,

400

250,

500

1000,

2000

2000,

4000

2000,

4000

Pan size (in / cm)

3.5 / 9 dia.

4.7 / 12 dia.

7.1 / 18 dia.

Net Weight (lb/kg)

10.1 / 4.6 (

*

11.2 / 5.2)

10 / 4.5 (

*

11.1 / 5.1)

CTS

RTS

5

9

1

6

RxD

TxD

GND