Replacement parts, Accessories, Specifications – Ohaus ADVENTURER BALANCES Manual multi User Manual
Page 9: Error codes

5
PARTS INFORMATION
If you require replacement parts or would like to purchase
accessories, inside the U.S., please call Ohaus Corporation
toll-free at (800) 526-0659 and from outside the U.S., (973)
377-9000. An Ohaus Product Parts Specialist will be available
to help you.
REPLACEMENT PARTS
Part No.
AC Adapters:
US/Japan 100 V
11103743
US 100-120V
11103741
Europe 220-240V
11103740
UK 220-240V
11103742
Table version 220-240V Reqs. sep. line cord, Aus. 11103745
Line cord for above
76199-01
Table version 220-240V Reqs. sep. line cord, US 11103745
Line cord for above
6569-00
Euro/Germany, Belgium, France 220-240V
11103744
Pan for 65g, 110g, 210g 3.5" (9cm) dia.
12106739
Pan for 150g, 310g,
3.9" (10cm) dia.
12105372
Pan for 510 to 4100g,
7.1" (18cm)
12105370
ACCESSORIES
Part No.
20g Calibration Mass
49024-11
50g Calibration Mass
49054-11
100g Calibration Mass
49015-11
200g Calibration Mass
49025-11
500g Calibration Mass
49055-11
1kg Calibration Mass
49016-11
2kg Calibration Mass
49026-11
Security Device
76288-01
Impact printer
SF42
In-Use Cover
400302-010
Error Codes
When a problem occurs using the balance, the display will indi-
cate an error code. Review the listed codes and follow instruc-
tions to correct the problem.
Err 2.0 Stability error. Check if balance is located near vibrat-
ing equipment or if air currents are affecting it.
Err 3.0 Calibration error (wrong mass). Incorrect or no cali-
bration mass used when performing calibration proce-
dure. Error will flash momentarily, then balance will
use previous calibration data. Recalibrate correctly.
Err 4
Invalid data checksum in type or adjust data. Return
balance for servicing.
Err 6.0 Parts counting error - Average piece weight is less than
1d. Balance shows error then exits parts counting.
Err 8.4 Over or under load. Sample being weighed exceeds
the capacity of the balance. If error occurs when the
sample is within the balance capacity, balance may be
incorrectly calibrated. An underload such as the pan
off of the balance could also display Err 8.4. Recali-
brate the balance.
Err 9
Internal data error. Return balance for servicing.
Err 9.8 Invalid checksum in calibration or setup data. The
balance may need recalibration, particularly linearity
calibration. If the error persists after recalibration,
the balance must be serviced.
Capacity
65g
(1)
110g
(1)
210g
(1)
150g
(1)
210g
(1)
310g
(1)
410g
(1)
510g
1500g 2100g
3100g 4100g
Readability
0.1mg 0.001g
0.01g
0.1g
Weighing units/modes
g, kg, mg, ct, N, lb, oz, ozt, gn, dwt, momme, mesghal, 3 taels, ticals, Parts Counting
Repeatability (Std. Dev.)
0.1mg
0.001g
0.01g
0.1g
Linearity
±0.2mg
±0.3mg
±0.002g
±0.01g ±0.02g
±0.1g
Tare Range
Full capacity by subtraction
Stabilization Time (seconds)
3 Seconds
Sensitivity Drift
4ppm/°C
10ppm/°C
Operating Temp Range
50° to 86° F/10° to 30° C
Power Requirements
8 - 14.5V 50/60Hz 6VA or 9.5 - 20V 6W
Calibration
External Digital
Display (in/cm)
LCD (1.2/3 high)
Pan Size (in/cm)
3.5/9 dia.
3.9/10 dia.
7.1/18 dia.
Draft shield clearance height
9.3/23.6
6.8/17.25
above pan (in/cm)
Dimensions (WxHxD) (in/cm)
8.5x14.3x13.5/21.7x36.3x34.3 8.5x12.3x13.5/21.7x31x34.3
8.5x4.3x13.5/21.7x11x34.3
Net Wt (lb/kg)
12.5/5.7
10/4.5
8.5/3.9
Shipping Wt (lb/kg)
21/9.5
16/7.3
14/6.4
Item No.
AR0640 AR1140 AR2140 AR1530 AR2130 AR3130 AR4130 AR5120 ARA520 ARB120 ARC120 ARD110
SPECIFICATIONS
Notes 1. 65g to 310g analytical balances include a three door draft shield.
2. Some models are not available in all countries.