5

Following the instructions provided with the indicator, wire the load cell cable to the
terminal block as shown in the table below.

WIRE COLOR

DESIGNATION

Green

+Excitation

Black

- Excitation

Red

+Signal

White

- Signal

Braided Shield

Ground

Yellow

Unused

Blue

Unused

Be sure to insulate the braided shield and the unused yellow and blue wires to
prevent electrical shorting to the indicator's circuit board.

Connection to non Ohaus Indicators

Connection to non Ohaus Indicators

Connection to non Ohaus Indicators

Connection to non Ohaus Indicators

Connection to non Ohaus Indicators

Connections should be made according to the instructions provided with the
indicator.