EN-4 Valor 3000 Series Scales

Figure 2-3. Power Connection and Sealing

.

2.5 Initial Calibration

When the scale is first installed, it should be calibrated to ensure accurate weighing
results.
Press and hold the ENTER/Cal button until CAL is displayed. When the button is
again pressed and released, the display shows -C- while zero reading is stored.
Next, the display shows the calibration weight value in grams. Place the specified
calibration weight (sold separately) on the pan. See Table 2-1 for appropriate
weights. The display shows -C- while the reading is stored. The display returns to
the previous application mode and scale is ready for use.

TABLE 2-1. CALIBRATION WEIGHTS.

MODELS

V31XH202

V31XH402

V31XH2 V31XH4 V31X501

Calibration
Weights

200g

400 g

2 kg

4 kg

500 g

MODELS

V31X3

V31X6

V31XW301

V31XW3 / V31X3N

V31XW6 / V31X6N

Calibration
Weights

3 kg

6 kg

300 g

3 kg

6 kg

3. OPERATION

3.1 Weighing Mode

Use this mode to measure the weight of a sample in the selected unit of measure.

1. Press

MODE/UNIT/Menu until the desired unit icon is displayed.

2. Put an empty container on the pan, then press O/T/On/Off to tare the container
weight.
3. Add material to the container. The display shows the weight of the material.