6. MAINTENANCE

6.1 Troubleshooting

The following table lists common problems and possible causes and remedies.
If the problem persists, contact Ohaus or your authorized dealer.

TABLE 6-1. TROUBLESHOOTING.

Symptom

Possible Cause

Remedy

Cannot turn on

No power to scale

Verify connections and voltage.

Poor accuracy

Improper calibration • Unstable
environment

Perform calibration • Move scale to
suitable location

Cannot calibrate

Unstable environment •
Incorrect calibration weight

Move the scale to suitable location
• Use correct calibration weight

Cannot access
mode

Mode not enabled

Enter menu and enable mode

Cannot access unit Unit not enabled

Enter menu and enable unit

Err 8.1

Pan has load during power on

Remove weight from pan and re-
zero.

Err 8.2

Pan was removed prior to
power on

Install pan and re-zero.

OVER

Weight on pan exceeds
capacity

Remove weight from the pan

UNDER

Pan was removed during
weighing

Re-install pan

Err 9

Factory calibration data
corrupted

Contact Ohaus or an authorized
service agent • Perform service
calibration

REF Err

Reference Weight is too small Use larger sample

LOWrEF

Reference Weight is too low for
accurate percent weighing.

Continue to weigh with less
accurate results

6.2 Service Information

If the troubleshooting section does not resolve or describe your problem, contact
your authorized Ohaus service agent. For service assistance or technical support in
the United States call toll-free 1-800-526-0659 between 8.00 AM and 5:00 PM
EST. An Ohaus product service specialist will be available to provide assistance.
Outside the USA, please visit our web site, www.ohaus.com to locate the Ohaus
office nearest you.