Basic Description

Specifications

2

Additional Alarms and Indicators

©2011 General Electric Company – All rights reserved.
GE and GE Monogram are trademarks of General Electric Company.
*iVent, Easy Exhale, Adaptive Bi-Level, Adaptive Flow and Adaptive I-Time
are trademarks of General Electric Company.
VersaMed, a General Electric Company, doing business as GE Healthcare.
General Electric Company reserves the right to make changes in specifications and features
shown herein, or discontinue the product described at any time without notice or obligation.
Contact your GE Representative for the most current information.
CAUTION: USA Federal law restricts this device to sale by or on the order of
a licensed physician.


DOC0982719 Rev.1

2. Some ventilation modes and features are optional and may vary according to the
selected model.

GE Healthcare
P.O. Box 7550
Madison, WI 53707-7550, U.S.A.
Tel 800-345-2700
www.gehealthcare.com/RespiratoryCare

Ventilation Performance and Controlled Parameters

Respiratory rate:

1 to 80 bpm

Tidal volume:

50 to 2,000 ml

Inspiratory pressure limit:

5 to 80 cm H

2

O

Inspiratory time:

Adaptive I-Time* or 0.2 to 3 sec

Esens:

10-90%

Rise time:

Mid/high/max/auto

Peak flow:

Adaptive Flow or 1 to 120 L/min

Spontaneous up to 180 L/min

Oxygen mix (FIO

2

):

21 to 100%

PEEP:

0 to 40 cm H

2

O

Trigger sensitivity:

1 to 20 L/min flow sensitivity

–0.5 to –20 cm H

2

O pressure sensitivity

PSV:

0 to 60 cm H

2

O

Positive pressure relief ralve:

80 cm H

2

O

Synchronized nebulizer:

5 to 240 min

Monitoring and Displayed Parameters

• Airway pressure

(analog bar graph & numerical)

• Total breath rate
• I:E ratio
• Exhaled tidal volume
• Exhaled minute volume
• Peak flow
• Inspiratory time
• Electrical power source

(AC/external DC/internal DC)

• Battery level
• Pressure, flow and volume Waveforms

software package:

– Real-time pressure and

flow waveforms

– Waveform history browse
• Trending of monitored data (72 hours)
• Respiratory mechanics (C, R, MAP, RR/Vt)
– Pressure, flow and volume loops
• SPO

2

User Adjustable Alarms

Alarms:

• Inverse I:E ratio

• Low O

2

pressure

• AC disconnect

• Low/Empty battery

• Over temperature

• Service notice

• Patient disconnect

• Check sensor

• High PEEP
• Battery disconnected / damaged
• Sensor disconnect

Indicators:
• Alarm silence icon and timer
• Breath type icon
• Internal battery use
• Date and time
• Hour meter
• Battery charge level
• Need calibration
• LED: On, charge, alarm
• Nebulizer status indicator

Size and Weight

Dimensions:
Height:

13”/33 cm

Width:

9.5”/24 cm

Depth:

10.3”/26 cm

Screen:

8.4” diagonal

Weight:

18.3 lb/8.3 kg (without battery)

Battery weight:

6.2 lb/2.8 kg

Overall weight with standard battery:

24.5 lb/11.1 kg

Overall weight with optional extended
internal battery:

27.8 lb/12.6 kg

Power Supply

External AC:

100 to 240 V, 50 to 60 Hz, Max 1.6 A

External DC:

12 to 15 V, Max 8.5 A

Internal battery:

Sealed lead-acid 12 V (7.8 Ah) (rechargeable)

Operating time:

Depending on ventilator settings and impedance –

standard internal up to 2 hours, optional extended internal
up to
4 hours and optional external up to 8 hours.

Oxygen (Enrichment) Supply

High pressure supply:

40 to 75 psi (2.8 to 5.1 bar)

Low pressure:

Max 15 L/min or 0.5 psi

External Interface

Environmental Specifications

Operating temperature: 0 to 50° C/32 to 120° F
Storage temperature:

–15 to 70° C/–4 to 140° F

Relative humidity:

15 to 95% at 30° C/85° F

Water/dust resistance: IP54 (splash proof)
Atmospheric pressure: 430 to 825 mm Hg (15,000 feet)
Vibration:

IEC 68-2-6 and IEC 68-2-34

MIL-STD-810E

Shock:

IEC 68-2-27 (100 g)

MIL-STD-810E

Total external sound level: 40 to 45 dBa at one meter

Standards and Safety Requirements

• Electrically powered,

computer controlled ventilator

• Closed loop flow & pressure controller
• Ventilation modes:
– Assist control (A/C):
– Volume controlled
– Pressure controlled
– Synchronized intermittent

mandatory ventilation (SIMV):

– Volume controlled
– Pressure controlled
– Continuous positive airway pressure

(CPAP)

– Pressure support ventilation (PSV)

• Special features:
– Adaptive Bi-Level* (NIV)
– Adaptive Flow*

– Adaptive I-Time*

– Preset parameters by patient weight
– Apnea backup ventilation
– Adjustable rise time
– Easy Exhale
– Sigh breath
– 100% O

2

suction

• Respiratory rate (high/low)
• Minute volume (high/low)
• Pressure (high/low)

• Low Vt (15 to 85%)

• SPO

2

(high/low)

• HR (high/low)
• Apnea time (0 to 120 seconds)
• FIO

2

(high/low)

• Leak (0 to 100%)

• Remote monitor (VGA)
• RS-232 serial port, 9 pin

• DIN keyboard connector
• RJ11 remote alarm connector

Meets the requirements of:
• ASTM F1246-91
• CSA C22.2 No. 601.1/601.2
• UL 60601.1
• EN/IEC 60601-1
• EN/IEC 60601-1-2

• EN/IEC 60601-2-12
• EN 794-3
• ISO 10651-2/3
• EN/IEC 60601-1-6
• EN/IEC 60601-1-8