Imagination at work, Clinical performance, Asystole alarms – GE Healthcare IntelliRate Technology for Patient Monitoring User Manual
Page 4: Bradycardia alarms, Tachycardia alarms, Conclusion, References, Additional resources

Clinical Performance
The IntelliRate software algorithm has been evaluated 
on 55 CICU patient recordings equaling over 900 hours 
of monitoring time. After reviewing the multi-parameter 
waveform data to positively identify all true Tachycardia, 
Bradycardia, and Asystole conditions, alarm performance 
was measured when only ECG information was used.
2
Next,
alarm performance was measured using the IntelliRate 
software to demonstrate the improvement in alarm accuracy. 
The results of the comparison appear in Figure 1.
Asystole Alarms
True Alarms
False Alarms
ECG Only
0
18
IntelliRate
0
2
“False Asystole Alarms Reduced by 89%”
Bradycardia Alarms
Sensitivity
Positive 
Predictive Value
ECG Only
96%
58%
IntelliRate
94% 84%
“False Low Heart Rate Alarms Reduced by 72%”
Tachycardia Alarms
Sensitivity
Positive 
Predictive Value
ECG Only
93%
86%
IntelliRate
93% 88%
“Findings not statistically signifi cant“
Figure 1. IntelliRate Performance
Conclusion
IntelliRate software represents a major advance in eff orts 
to reduce the occurrence of false alarms in critical care 
environments. By utilizing the signal information from 
multiple parameters, the accuracy and reliability of heart rate 
measurements can be signifi cantly improved. This enhanced 
accuracy, when used in the context of data available from 
other monitored parameters, signifi cantly reduces false 
alarm rates and can help enhance safety for pacemaker 
patients.
References
1 Tsien, C.L. and J.C. Fackler, Poor Prognosis for Existing
Monitors in the Intensive Care Unit, Critical Care Medicine 
25, 614-619 (1977).
2 The ECG alarm performance was established using GE
EK-Pro ECG analysis algorithm. This algorithm has been 
shown to be over 99% accurate in QRS detection as 
measured on the standardized MIT-BIH database.
imagination at work
©2010 General Electric Company – All rights reserved.
GE and GE Monogram are trademarks of 
General Electric Company.
IntelliRate is a trademark of GE Medical Information 
Technologies Systems, Inc.
General Electric Company reserves the right to make 
changes in specifi cation and features shown herein, or 
discontinue the product described at any time without 
notice or obligation. Contact your GE representative for 
the most current information
GE Healthcare, a division of General Electric Company
DOC0804579 8/10
GE Healthcare
8200 W. Tower Ave.
Milwaukee, WI 53223
USA
www.gehealthcare.com
GE Healthcare Finland Oy
Kuortaneenkatu 2
00510 Helsinki
Finland
GE Healthcare
3/F Building # 1, 
GE Technology Park
1 Hua Tuo Road
Shanghai 201203
China
Additional resources
For white papers, guides and other instructive materials about 
GE Healthcare’s clinical measurements, technologies and 
applications, please visit http://clinicalview.gehealthcare.com/