Technical specifications – GE Healthcare Vivo 40 User Manual

Page 2

background image

Technical Specifications

Settings/Performance
Ventilation Modes

• PSV (Pressure Support Ventilation)
• PCV (Pressure Control Ventilation)
• PSV +Target Volume
• PCV +Target Volume
• CPAP (Continuous Positive Airway Pressure)

Patient Modes

• Adult
• Paediatric (>10 kg)

Device Modes

• Clinical
• Home

IPAP/EPAP

4 to 40 cmH

2

O / 2 to 20 cmH

2

O or IPAP -2 cmH

2

O

CPAP

4 to 20 cmH

2

O

Breath Rate

4 to 40 BPM

Inspiratory Time

0.3 to 5 s

Rise Time

1 to 9

Inspiratory Trigger (eSync)

1 to 9, Off

Expiratory Trigger

1 to 9

Target Volume

200-1500 ml

Expiratory Trigger

1 to 9

Ramp function setting

10 – 60 min, Off (PCV, PSV and CPAP mode)

Humidification setting

1 to 9 (10 to 30 mgH

2

O/L)

Audible alarm setting

1 to 9

Maximum flow

> 200 litres/min

Minimal Inspiratory Time

Off, 0.3 – 3.0 s

Maximal Inspiratory Time

0.3 – 3.0 s, Off

Noise level (at 10 cmH

2

O constant pressure)

Less than 30 dB(A)

Indications

Pressure; bar graph and digital data

0 to 40 cmH

2

O

Calculated tidal volume accuracy

± 20%

Leakage calculation accuracy

± 20%

Total Breath Rate

BPM (Breaths Per Minute)

I/E Ratio

1 :10 to 10 :1

Power Supplies

Mains supply

100 to 240 V AC

Internal battery / External battery Capacity 3.8 Ah / 12/24 V DC

Operating Conditions

Recommended mask leakage

20 to 40 litres/min at 10 cmH

2

O

Minimum mask leakage

12 litres/min at 4 cmH

2

O

Alarms

Audible Alarm Level

45 to 85 dB(A)

Low pressure alarm, High pressure alarm, Low volume alarm, Low breath rate alarm,
High breath rate alarm, Low leakage alarm, High leakage alarm, Low External DC War-
ning, Low internal battery warning, Low power alarm, Internal function failure alarms.

Dimensions

W x H x D

185 x 240 x 227 mm

Weight

4.0 kg

Air outlet

22 mm male conical standard connector

Breas Medical AB · Företagsvägen 1 · SE-435 33 Mölnlycke · Sweden

Phone +46 31 86 88 00 · Fax +46 31 86 88 10 · [email protected] · www.breas.com · www.gehealthcare.com

* Vivo by Breas and eSync are trademarks by General Electric Company.

© 2012 General Electric Company – All rights reserved.

General Electric Company reserves the right to make changes in specifications and features

shown herein, or discontinue the product described at any time without notice or obligation.

Contact your GE representative for the most current information.