Technical specifications, Environmental specifications, Physical characteristics – GE Healthcare Stand-alone Resuscitation System User Manual

Page 4: System characteristics, Airway pressure manometer, Air/o, Blender, User control settings, Recommended body weight range, Approximate operating time

background image

© 2012 General Electric Company – All rights reserved.

General Electric Company reserves the right to make
changes in specifications and features shown herein,
or discontinue the product described at any time without
notice or obligation. Contact your GE Representative
for the most current information.

GE, the GE Monogram and imagination at work are
trademarks of General Electric Company.

* Giraffe and Panda are trademarks of General Electric
Company.

Datex-Ohmeda, Inc., a General Electric company
doing business as GE Healthcare.

Environmental specifications

O

perating temperature

18 to 41°C

Operating humidity

0 to 90%

Storage temperature

–25 to 60°C

Storage humidity

0 to 95%

Storage pressure

50 to 106 kPa

Physical characteristics

Depth (knobs to fittings)

<8.7 in (22 cm)

Height

<14.6 in (37 cm)

Width

<11.8 in (30 cm)

Weight

Approx. 15 lbs (7 Kg)

System characteristics

I

nput source

Hospital pipeline or

cylinder with regulator

Input pressure

40-75 psi (275-600 kPa)

(Air/O2 fittings)
Minimum input

70 lpm

Vacuum range

0-150 mm Hg

Vacuum gage accuracy

±5% of full scale

Flow capacity

15 lpm

Flow range

0-15 lpm

Flow accuracy

–5 lpm ±½ lpm

5-10 lpm ±1.5 lpm

10-15 lpm ±2 lpm

Airway pressure manometer

R

ange

–10 to 80 cmH

2

O

Accuracy

±5% of full scale

Air/O

2

blender

Range

21-100% O

2

Accuracy

±5% of full scale

User control settings

Air/O

2

flowmeter

0-15 lpm

set range
Suction regulator

0–150 mmHg

set range

Recommended body weight range

Up to 22 lbs (10kg)

Approximate operating time* *

@ 5 L/min

121 min

@ 10 L/min

74 min

@ 15 L/min

54 min

For T-Piece Resuscitation System only
Adjustable PIP range

(located on the resuscitation front panel)

Maximum PIP:

45 ±5 cmH

2

O

Override:

>30 ±4 cmH

2

O

Adjustable PEEP range

(located on the T-Piece circuit)

@ 5 L/min minimum PEEP

<5 cmH

2

O

@ 8 L/min minimum PEEP

<5 cmH

2

O

@ 10 L/min minimum PEEP

<5 cmH

2

O

@ 15 L/min minimum PEEP

<6 cmH

2

O

* *E cylinders (680 liters), suction off, 50% Air/O

2

Technical Specifications

EMEA DOC1106181 Rev2 7/12