5
19
Installation
Bipolar High-Stability Supply
f. Connect the negative magnet current lead (3) to the negative power
supply bus (6).
g. Connect Model 08150PS Master unit terminal block positions CS,
S-, and S+ to each corresponding Slave terminal (18). Connect
Master to first Slave, first Slave to second Slave, etc.
h. Connect two jumpers (17) from terminal block position S- to M- and
from S+ to M+ on the Master
1
power supply unit.
i. Connect the special DB15 analog I/O cable from the
PROGRAM
Connect the CT power supply to the
TRANSDUCER POWER
connector (5) on the rear of the Model 430 Programmer.
j. Connect the current transducer output to the current transducer
SIGNAL
connector (4) on the rear of the Model 430 Programmer.
k. Connect the special DB15 analog I/O cable from the
PROGRAM
OUT
connector (14) on the back of the Model 430 Programmer to
the DB15
ANALOG I/O
connectors on the rear of the Master (7) and
each Slave (20) power supply unit.
l. Install an instrumentation cable between the magnet support
stand top plate connector (8) and one of the
MAGNET STATION
connectors (11) on the rear of the Model 430 Programmer.
m. Optional: Install an instrumentation cable between one of the
LHe /
TEMP
connectors (10) on the rear of the Model 430 Programmer
and the Model 13x Liquid Helium Level Instrument and/or
temperature instrument (9). Refer to section A.2 on page 152.
n. Optional: Install an instrumentation cable between the
QUENCH
I/O
connector (13) on the rear of the Model 430 Programmer and
Aux connector
J2
on the rear panel of the Model 13x Liquid Helium
Level Instrument (12). Refer to section A.6.2 on page 157.
o. Connect each device line cord and Model 601 power adapter from
the respective device to the appropriate power receptacle.
p. Remote communications via Ethernet and/or RS-232 can be
accomplished by connecting suitable cabling to the Model 430
Programmer rear panel
ETHERNET
and/or
RS-232
connectors.
1. Do not add terminal block jumpers to the slave units. Only the single or master unit
is jumpered.