ReCon TM Magnet Systems
Based on heat load and size of the superconducting magnet
either a dry conduction cooled or ReCon
TM
style geom-
etry is recommended. ReCon
TM
style magnet systems use
cooling power of a closed cycle 4K cryocooler to recon-
dense gaseous helium back into the liquid Dewar. The
evaporating gas is liquefied in a thermodynamically ef-
fective manner and the cryocooler is strategically located
in the Dewar. The first stage of the cryocooler is used to
cool the shields of the Dewar thereby eliminating the need
for any liquid Nitrogen. ReCon
TM
magnet systems offer
combined benefits of traditional LHe systems with near
zero helium boil-off. These style systems are ideally suited
for retrofitting existing magnets and sample inserts into
a re-engineered Dewar with a suitable gas handling sys-
tem. Pulse tube cryocoolers are ideal for applications that
desire lower vibrations.
200mm RTB ReCon system
ReCon
•
TM
solenoid systems available in vertical or horizontal geometries with
fi elds up to 16T and wide bore sizes
ReCon Maxes
•
TM
systems are available with vector fi elds up to 6T
ReCon OptiMAxes
•
TM
systems are available in 2 or 3-axis geometries
ReCon
•
TM
split coil systems available in optical and non-optical confi gurations
with fi elds up to 10T
ReCon
•
TM
systems with customized gas handling system is available with 1.5K/
He3 insert or with a dilution refrigerator