Activeforever Inogen One G3 Portable Oxygen Concentrator User Manual
Page 45

41
Inogen One® G3 System Specifications Chapter 8
English
ELECTrOMAGNETIC COMPATIBILITY
This CE Marked equipment has been tested and found to comply with the EMC limits for
the Medical Device Directive 93/42/EEC [EN 55011 Class B and EN 60601-1-2]. These limits
are designed to provide reasonable protection against harmful interference in a typical
medical installation.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Inogen One® G3 Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The customer or the user of the Inogen One® G3 Oxygen
Concentrator should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The Inogen One® G3 Oxygen
Concentrator uses RF energy only for
its internal function. Therefore its RF
emissions are very low and not likely
to cause any interference in nearby
equipment.
RF emissions
CISPR 11
Class B
The Inogen One® G3 Oxygen
Concentrator is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies