Electromagnetic immunity IEC60601-1-2
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3	V	rms
150	kHz	~	
80 MHz
3	V/m
80 MHz to
2.5 GHz
3 V rms
3 V/m
Portable and mobile RF 
communications equipment should 
be used no closer to any part of 
the NE-C801 including cables, 
than the recommended separation 
distance calculated from the equation 
appropriate to the frequency of the 
transmitter.
Recommend separation distance
d	=	1.2	√P
d	=	1.2	√P 80 MHz to 800 MHz
d	=	2.3	√P 800 MHz to 2.5 GHz
where P is the maximum output 
power rating of the transmitter 
in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in 
meters (m).
Field strengths from fixed RF 
transmitters as determined by an 
electromagnetic site survey,*
2)
should
be less than the compliance level in 
each frequency range.*
3)
Interference may occur in the 
vicinity of equipment marked with he 
following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects, and people.
*
2)
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless)
telephones	and	land	mobile	radio,	AM	and	FM	radio	broadcast,	and	TV	broadcast	cannot	be	
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed 
RF transmitters, an electromagnetic site survey should be considered. If the measured field 
strength in the location in which the NE-C801 is used exceeds the applicable RF compliance 
level above, the NE-C801 should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or 
relocating the NE-C801.
*
3)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
GUIdANCE ANd MANUfACTURER'S dECLARATION