19

dAILy dISINfECTING

Disinfect the nebulizer kit and mouthpiece, or optional masks or optional
nosepiece after the last treatment of the day.
To disinfect the parts, use one of the methods described below:
A. Use a commercially available disinfectant. Follow the instructions provided

by the disinfectant manufacturer.

1

Submerge the parts in the cleansing solution for the specified period.

2

Remove the parts and discard the solution.

3

Rinse the parts with clean hot tap water, shake off excess water and allow

to air dry in a clean environment.

B. Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake off excess water and allow to
air dry in a clean environment.

nOTe: Do not boil air tube, air filter, air filter cover and optional masks.

CaUTIOn

Do not use an autoclave, EOG gas disinfection or low temperature plasma sterilizer to

disinfect the device.

handling the Baffle

The baffle is an important part that is used to nebulize medication.
Be sure to observe the following precautions when handling it.

CaUTIOn

• Always wash the baffle after each use.
• Do not use a brush or pin etc., to clean the parts.
• When disinfecting the parts by boiling, be sure to

boil them in plenty of water.

• Do not boil the baffle together with other objects,

other than applicable nebulizer accessories.

• Do not use a microwave oven, dish dryer or hair

dryer to dry parts.