SECTION 7—CLEANING, CARE, AND MAINTENANCE
Invacare SOLO
2
™
32
Part No 1156872
SECTION 7—CLEANING, CARE, AND
MAINTENANCE
ƽ WARNING
Turn Off the transportable oxygen concentrator and unplug the power
cord before cleaning.
DO NOT allow any cleaning agent to drip inside the air inlet and outlet
openings, or the battery pack.
DO NOT spray or apply any cleaning agent directly to the cabinet.
CAUTION
DO NOT clean the cabinet, carry bag, or filter with alcohol and alcohol
based products (isopropyl alcohol), concentrated chlorine‐based products
(ethylene chloride), and oil‐based products (Pine‐Sol
®
, Lestoil
®
) or any
other harsh chemical agents. Only use mild liquid dish detergent (such as
Dawn
™
).
Cabinet
Periodically clean the transportable oxygen concentrator cabinet when
dirty as follows:
1.
Use a damp cloth, or sponge, with a mild detergent such as Dawn dish
washing soap to gently clean the exterior case.
2.
Allow the unit to air dry, or use a dry towel, before operating the unit.
Air Intake Filter
CAUTION
DO NOT operate the transportable oxygen concentrator without the air
intake filter installed.
NOTE: For this procedure, refer to FIGURE 7.1.
1.
Clean the air intake filter at least once a week depending on
environmental conditions.
2.
Squeeze the thumb tabs on the inlet filter grill and remove from the
unit.
3.
Lift out the filter.
4.
Use a vacuum cleaner or wash with a mild liquid dish detergent (such
as Dawn™) and water. Rinse thoroughly.