beautypg.com

2 flow rate indicated is not correct – Toshiba LF414 User Manual

Page 42

background image


- 41 -

6F8A0870

7.2.2 Flow rate indicated is not correct


























































Is the flow range correctly
set?

Set correctly.
Refer to combined converter's manual.

NO

YES

Is zero point correctly set?

Perform

the

zero

adjustment.

Refer to combined converter's manual.

NO

YES

Is

the

excitation

current

value as stated on the
flow direction tag?

Set correctly.
Refer to combined converter's manual.

NO

YES

Is

the

inside

wall

of

detector

pipe contaminated?

Clean the inside of the detector pipe.

YES

NO

Is

the

output

load

resistance

as specified?

Design the output circuit so that

the

output

load resistance is less than

750

Ω

.

NO

YES

Calculate as shown on the left.

Are there two load

resistors connected to

the output in parallel?

Connect the two load resistors in
series, if necessary.

YES

NO

NO

YES

Contact your nearest Toshiba
representative.

START

Is accuracy calculated as follows?

(Measured flow rate)-(Actual flow rate)

Actual flow rate

Ч

100%