beautypg.com

18 error displays – Topcon GTS-823A User Manual

Page 129

background image

18-1

18 ERROR DISPLAYS

18

ERROR DISPLAYS

l

If error still persist after attempting to clear them, contact your local Topcon dealer or Topcon head
office.

Error code

Description

Countermeasures

Backup

battery

empty

Displayed when built in battery for memory
back up is empty.

Contact your dealer or Topcon.

W/C OVER

Displayed when measurement carried out
within ±9

°

from zenith or nadir at the Earth

curvature and refraction correction mode is
ON.

Set correction for refraction and earth
curvature mode OFF or measure out of ±9

°

from the zenith or Nadir.

H angle

measuring

error

Displayed when the instrument rotated too
fast or any abnormality occurs in angle
measuring system .

The instrument will return to previous mode
automatically.

V angle

measuring

error

Displayed when the telescope rotated too fast
or any abnormality occurs in angle measuring
system.

The instrument will return to previous mode
automatically.

E31

Displayed when the unit of the angle at the
recall mode is different from the unit stored in
setting mode.

Make the unit in same unit system.

E35

Displayed when REM measurement carried
out to the range from zenith or nadir ± 6

°

.

Operate in the range out of ± 6

°

from the

zenith or nadir.

E36

Displayed when the N.E coordinates are set
same as the instrument coordinate in setting
direction angle or lay out mode.

Set except the instrument coordinate value.

E60’s

Any abnormality occurs with EDM (distance
measuring system).

Repair is required.

E71

Displayed when Vertical angle 0 position is
set with incorrect procedure.

Confirm the procedure and readjust.

E72

Displayed when Vertical angle -position is
adjusted in wrong position.

Confirm the procedure and readjust

E73

The instrument was not leveled when Vertical
angle 0-position is adjusted.

Level the instrument then carry the
adjustment work.

E81
E82

Mainly at the time data transmission between
GTS-820 series and external instrument.

Press [F1](EXIT) key, and confirm the
connection cables are correct.

Other
E80‘s

Data transmission error between internal
P.C.B.‘s.

Restart and confirm the operation procedure
is correct.

E90‘s

Abnormality in internal memory system.

Repair is required.

E300‘s

Abnormality in tracking system.

Repair is required.

E600‘s

Abnormality in angle measuring system.

If this error code continues to display, repair is
required.

E700‘s

Abnormality in angle measuring system.

If this error code continues to display, repair is
required.

E800‘s

In the self checking mode, vibration is too big. Get rid of the Vibration

This manual is related to the following products: