Table 8: Error Codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a
numeric entry

101

Too many digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid command parameter was received

107

Pressure not selected

108

Invalid sensor type

109

Pressure module not connected

110

An unknown command was received

111

Bad Parameter received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

10.6 Entering Commands

Commands for the calibrator may be entered in upper or lower case.
There is at least one space required between the command and
parameter, all other spaces are optional. Almost all commands for
the calibrator are sequential; any overlapped commands will be
indicated as such. This section will briefly explain each of the
commands and describe their general use, which will include any
parameters that may be entered with the command as well as what
the output of the command is.

10.6.1 Common Commands

*CLS

Clears the error queue. Also terminates all pending operations. When
writing programs, use before each procedure to avoid buffer
overflow.

*IDN?

Returns the manufacturer, model number, and firmware revision of
the Calibrator. For example:

*IDN? will return OMEGA, PCL340, 0, 1.00