Functions setting

(1) Inflation level setting

AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure
(approximately 30-40 mmHg above the patient’s systolic pressure).

Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the
level to 30 to 40 mmHg higher than the expected systolic pressure.

• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.

• In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than

220 mmHg. Use the unit in the manual setting.

• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.

ON/OFF

AVG /1st/2nd/3rd

DEFLATION

START

STOP

HIDE

P-SET

100

AUTO

140

180

220

260

280

MODE

AVG.

MANU.

CHECK

SINGLE

ON/OFF

AVG /1st/2nd/3rd

DEFLATION

START

STOP

HIDE

P-SET

100

AUTO

140

180

220

260

280

MODE

AVG.

MANU.

CHECK

SINGLE

(2) Non-display function

Use to prohibit the display of measurement results. However,
the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.

By pushing the

Button, display or non-display of status

is switched alternatingly.

HIDE

(3) Manual deflation control

Accelerate deflation by pushing the DEFLATION (deflation
control) / Measurement Result Display Switch Button during
deflation in the measurement by the MANU Mode.

With each push of the

Button, cuff is deflated rapidly

in increments of 5 to 10 mmHg.

BPM

mmHg

PULSE

SYS

mmHg

DIA

AC

AC

Display

Non-display

P-SET

HIDE

ON/OFF

AVG /1st/2nd/3rd

DEFLATION

START

STOP

HIDE

AC

P-SET

100

AUTO

140

180

220

260

280

MODE

AVG.

MANU.

CHECK

SINGLE

XL

ON/OFF

AVG /1st/2nd/3rd

DEFLATION

START

STOP

HIDE

AC

P-SET

100

AUTO

140

180

220

260

280

MODE

AVG.

MANU.

CHECK

SINGLE

XL

ON/OFF

AVG /1st/2nd/3rd

DEFLATION

START

STOP

HIDE

AC

P-SET

100

AUTO

140

180

220

260

280

MODE

AVG.

MANU.

CHECK

SINGLE

XL

List of Functions for each Mode

Inflation level set-

ting

Non-display func-

tion

Manual deflation

control

Single (SINGLE)

Mode

Average

(AVG) Mode

Auscultation

(MANU) Mode

Check

(CHECK) Mode

Function

Measurement

Mode

(4) AVG Function setting

You can set the number of measurements, the waiting time until the 1st measurement, and the
measurement interval for the AVG Mode.

Items to set

Set value

F1

Number of measurements

2 times or 3 times

F2

Waiting time until the start of 1st measurement

0 sec, 3 min, 5 min, or 10 min.

F3

Measurement interval

5 sec, 30 sec, 1 min, or 2 min.

Note: The bold letters represent the factory-set values.

Procedure to change the set values

1) When the power is OFF, push the ON/OFF (power) Button

for more than three seconds while holding the START Button;
F1 is displayed.

2) Push the START Button and select the function to set from

F1 to F3. Each time you push the START Button, the func-
tions change in the order of

.

3) Push the DEFLATION (deflation control)/Measurement

Result Display Switch Button and change the set values.

4) When the setting is finished, push the ON/OFF (power)

Button to turn off the power. The setting is changed.

min

sec

Function #

HEM-907 10/23/03 3:13 PM Page 12