HOW TO CLEAN THE UNIT AFTER USE

XL

1. Wipe the monitor with a soft, damp cloth diluted with disinfectant alcohol, or

diluted detergent.

2. Complete cleaning by wiping the monitor with a soft, dry cloth.

General advice

• Do not clean this unit with gasoline, paint thinner, or

high concentration alcohol.

• Do not disinfect this unit by autoclave or gas steril-

ization (EtO, gluteraldehyde, or high concentration
ozone.)

When cleaning this unit, unplug the AC
adapter from the electric outlet.

After cleaning this unit, dry it well, before
plugging the AC adapter in the electric outlet.

Accuracy of pressure display can be checked in the CHECK Mode.

What you need to prepare

(1) Calibrated mercury manometer (including inflation bulb), (2) T-tube, (3) two air tubes, and (4)
a sturdy cylindrical shaped object on which the cuff is wrapped.

Check result

Accuracy of the monitor is validated to be ±3 mmHg or 2% of standard manometer reading. If your
result shows a difference exceeding the tolerance, contact Omron repair department (1-877-216-1336).

How to check

1. Connect the manometer,

inflation bulb, cuff, and the
monitor with the T-tube as
shown in the figure on
the right.

2. Tightly wrap the cuff over a

sturdy cylinder.

3. Release the valve of inflation

bulb to remove the air inside
the cuff completely.

4. Push the ON/OFF (power) Button to turn on the monitor.

5. Set the MODE Selector to “CHECK”.

6. Close the valve of inflation bulb and inflate the cuff to the

pressure to be checked, based on the manometer read.

7. Compare the pressure values displayed on the monitor to

the one on the manometer.

HOW TO CHECK

PRESSURE ACCURACY