beautypg.com

Product information, Specifications, 1 specifications – Omega Speaker Systems 8 Channel Thermocouple USB Data Acquisition Module TC-08 User Manual

Page 9

background image

Product information

5

2

Product information

2.1

Specifications

Resolution

20 bits (16.25 NFR)

Thermocouple types

B,E,J,K,N,R,S,T

Number of input channels

8

Conversion time - per active channel

100 ms

Conversion time - CJC

100 ms

Uncalibrated accuracy

The sum of ±0.2% and ±0.5°C

Full scale input

±70 mV

Common mode range

±7.5 V

Overvoltage protection

±30 V

Input impedance

2 M

W

Input connectors

8 x miniature thermocouple

Output connector

USB upstream connector

Power requirements

No power supply required

Environmental conditions

0 to 50°C, 25% to 75% humidity
Quoted accuracy over 20 - 30°C
NOT water-resistant

Note: The resolution and accuracy depend upon the thermocouple type and the
temperature range. Below is a list of the thermocouples and temperature ranges
supported by the USB TC-08.

Actual measurable range

Theoretical ranges

CJC

-5°C to 70°C

N/A

Type B

20°C to 1820°C (CJC must be
>20°C)

0°C to 1820°C

Type E

-270°C to 910°C

-270°C to 1000°C

Type J

-210°C to 1200°C

-210°C to 1200°C

Type K

-270°C to 1370°C

-270°C to 1370°C

Type N

-270°C to 1300°C

-270°C to 1300°C

Type R

-50°C to 1760°C

-50°C to 1760°C

Type S

-50°C to 1760°C

-50°C to 1760°C

Type T

-270°C to 400°C

-270°C to 400°C