beautypg.com

Usb_tc08_set_mains – Omega Speaker Systems 8 Channel Thermocouple USB Data Acquisition Module TC-08 User Manual

Page 27

background image

Technical reference

23

3.3.2.6

usb_tc08_set_mains

short usb_tc08_set_mains
(

short handle,
short sixty_hertz

)

This routine sets the USB TC-08 to reject either 50 or 60 Hz.

Arguments

handle

Specifies the USB TC-08 unit.

sixty_hertz

Specifies whether to reject 50 Hz or 60 Hz. If set to 1,
the unit will reject 60 Hz, if set to 0, the unit will reject
50 Hz.

Returns

0

Use

usb_tc08_get_last_error

.

1

Mains rejection set correctly.

Note: If the rejection is not set correctly the unit will be more susceptible to mains
interference.