2.5

Clock Selection

The DAQ-12 is equipped with a programmable clock circuit to produce data sampling rates
independent from the clock rate of the host computer. An onboard 8254 programmable
interval timer, with a 10 MHz clock input and either two or three cascaded 16-bit timers,
provides the sampling rate. This enables the sampling rate to be adjusted from 5 us between
samples to almost a year between samples, in as small as 100ns increments.

The DAQ-12's sampling rate can also be generated from an external clock input. This external
clock can be connected directly to the A/D converter or through a 16-bit pre-divider, the
multi-function timer. Samples are taken on the low to high transition of the clock.

WARNING: For the DAQ-12, the maximum data sampling rate is 5us. This restricts

clock frequency to a maximum of 200 KHz. Sampling rates in excess of 200 KHz may

result in erratic operation and unpredictable results.

The clock source, internal or external clock, is software selectable through the DAQ-12's
control word register. The configuration of the clock source itself is controlled by jumper block
J3 as shown in Figure 2-6, (* indicates factory default).

5 6 7 8

1 2 3 4

J3

Internal Timer: 2 timers cascaded

connect 1-2, 4-8, 6-7*

3 timers cascaded

connect 1-5, 2-6, 7-8

External Timer: w/o pre-divider

connect 1-2, 3-7, 4-8

with pre-divider

connect 1-5, 2-3, 7-8

Figure 2-6. Jumper J3 Configuration

DAQ-12 Users Manual 17