Troubleshooting Guide

6

6-2

Problem

Solution

The display is either

a.

Clean the thermometer lens.

erratic or stays at

Refer to Section 5.1.

one reading

b.

Activate the Diagnostic Program in
the thermometer as follows:

c.

Press the

key.

d.

Hold down the

key and press

the

key until “VER X.X” is

displayed.

You can expect to see and hear the following:

• You will see the version number “VER X.X” of the

software for about 1 second.

• You will hear a beep, “TST” is displayed, and °F

flashes.

• Soon after, all of the segments of the display

including the backlighting will light up for about
1 second.

• The display will clear and a PAS (pass) or ERR (error)

code may be seen on the display.
If you see an error code, either “ERR1”, “ERR2”, or

ERR3”, record the code and call our Customer

Service Department. Provide Customer Service with
the error code that is displayed in the upper left
corner of the display. The Customer Service
Department representative may ask you to return the
thermometer to the factory.

• The display will go back to the Real Time Mode

(Emissivity Display Mode).

e. After running the diagnostic

program, press the

key to

release this mode.