3.6.1 Reviewing the Last Parameters

The thermometer stores the last temperature measured in

the real time mode (refer to Table 3-1). This temperature

can be recalled by pressing the

key.

1. With power applied to the unit, place the SLEEP/ON

switch to the SLEEP position.

2. Press the

key to review the most recently stored

temperature data and parameters. You may review:

• MAX temperature
• MIN temperature
• dIF temperature
• AVG temperature

• HAL temperature
• LAL temperature
• AMB temperature
• MEM location

3.6.2 Downloading Previously Stored Temperature Data

(OS553, OS554)

Up to 100 sets of stored temperature information can be
downloaded to a serial printer or a personal computer. Each set
of temperature information is broken down to the temperature
value, the Emissivity, and the high alarm setpoint for that
temperature.

1a. Turn on the Serial Printer and set the following

parameters:

Speed: 4800 BPS
Data:

8 Bits

One Stop Bit
No Parity

or

1b. Turn on the Personal Computer.

2.

Connect the RS-232 cable between the thermometer and
the Serial Printer or Personal Computer as shown in
Figure 3-10 or Figure 3-11.