Using the infrared thermometer – Omega Speaker Systems OS550 Series User Manual

Page 17

background image

3-1

Using the Infrared Thermometer

3

3.1 Using the Infrared Thermometer

3.1.1 Water Cool Jacket Accessory

When using the OS550 sensor head in an ambient temperature environment

above 50°C (122°F), the OS550-WC Water Cooling Jacket option must be

used to maintain the accuracy and response time of the unit. Two

1

8

"

N.P.T.

compression fittings are provided for connection to copper water lines. A

constant flow of approx. 0.5 GPM of clean, room temperature water is

sufficient to protect the instrument and maintain accuracy up to 85°C

(185°F). This option can be installed in the field.

Fig 3-1. Water Cool Jacket Dimensions

OS550-WC

Ш 70.4 (2.77)

63.5 (2.50)