beautypg.com

Using the down, Will cause – Omega Vehicle Security FPU5-MT-110 User Manual

Page 37

background image

Operating the Pump Motor in the STOPPED Display Mode

5

5-10

xxx mL/min

FLOW

RATE

ACTION

Displays current

FLOW RATE setting

Display shows

Using the UP ARROW key (ᮡ) will cause the FLOW RATE value to scroll up to
the maximum FLOW RATE determined by TUBE ID then scrolls over to the
minimum FLOW RATE.

Using the DOWN

ARROW

key

(

ᮢ)

will cause

the FLOW RATE value to scroll

down to the minimum FLOW RATE determined by TUBE ID then scrolls over
to the maximum FLOW RATE.

OR

xxx mL or xxx Liters

TOTAL

VOLUME

ACTION

Displays current

TOTAL VOLUME

setting

Display shows

Using the UP ARROW key (ᮡ) will cause the display to scroll up to a maximum of
10000 liters then over to 1 mL.

Using the DOWN ARROW key (

ᮢ) will cause the display to scroll down to 1 mL

then roll over to 10000 liters.

OR

xx:xx:xx

TIME

ACTION

Displays current

TIME setting

Display shows

Using the UP ARROW key (ᮡ) will cause the flashing TIME setting to scroll up
to maximum then roll over to “0”. The maximum time is 12 hours, 59 minutes
and 59 seconds (12:59:59).

Using the DOWN ARROW key (

ᮢ) will cause the flashing TIME setting to scroll

down to “0” then roll over to the maximum. The maximum time is 12 hours,
59 minutes and 59 seconds (12:59:59).

OR

The flashing group
(SECONDS, MINUTES, or
HOURS) is the one that will
scroll when the UP or
DOWN ARROW key
is
pressed. Pressing the TIME
key
again will select a
different group.

xxx rpm

RPM

ACTION

Displays current

RPM setting

Display shows

Using the UP ARROW key (ᮡ) will cause the RPM setting to scroll up to 600 rpm
then roll over to 10 rpm.

Using the DOWN ARROW key (

ᮢ) will cause the RPM setting to scroll down to

10 rpm then roll over to 600 rpm.

OR