Specifications
7
7-2
Average Temperature
Accuracy Time Period
30 days
(under continuous operation):
Emissivity:
0.10 to 1.00 in 0.01 increments,
set via keypad
Calculated Temperature
Maximum (MAX), Minimum (MIN),
Values:
Average (AVG), Differential (dIF),
Thermocouple (TC)
Ambient Target
Set and enabled via keypad
Temperature Compensation:(OS533E, OS534E, OS523E, OS524E)
RS232 Output (for
Standard on OS533E, OS534E, OS523E, OS524E
personal computers
9600 bits per second, 8 bits of data,
and serial printers):
1 stop bit, no parity
RS-232 Cable:
RJ12 to 9 pin D connector, Female
RJ12 Pin #
9 pin D connector Pin #
3 TX
2 RX
4 RX
3 TX
5 GND
5 GND
Analog Cable:
6 feet long; 2-conductor, 22 AWG
3.5mm male plug
Alarm:
Set and enabled via keypad
All models:
High alarm standard, with
audible and visual indication
OS533E, OS534E Low alarm standard, with
OS523E, OS524E: audible and visual indication
Data Storage:
OS534E, OS523E Up to 800 temperature data
OS524E:
points.
Aiming Feature:
”V” groove on top of the thermometer or use Laser
Sighting
Analog Output:
1 mV/°F or 1 mV/°C, set via keypad
(0.5 mV/Deg, OS524E)
Analog Output Accuracy: ±2mV reference to temperature display
Power:
4 AA size 1.5 volt batteries (lithium or alkaline)
Battery Type
Alkaline
General brand
Lithium:
Eveready Energizer, model number L91 BP-2
Battery Storage
–40°C to 50°C (-40°F to 122°F)
Temperature