Using the Handheld Infrared Thermometer

2

2-26

String

Description

E:95;

Emissivity is 0.95

MAX:78;

Maximum temperature is 78

MIN:65;

Minimum temperature is 65

DIF:13;

Differential temperature is 13

AVG:72;

Average temperature is 72

DIS:1144;

Distance is 11.44 feet (always in feet)

HAL:900;

High alarm set point (HAL) is 900

TC:74;

Thermocouple temperature is 74

TEF:0;

Thermocouple temp over-range flag (0: In range, 1: Out of range)

LAL:20;

Low alarm set point (LAL) is 20

AMB:125;

Target ambient temperature is 125

PRN:5;

Data transmission interval is every 5 seconds

PRNF:1;

PRN Flag (0: PRN disabled, No data transmission, 1: PRN enabled,
Data communication active)

IR:73;

Current Infrared temperature is 73

CF:0;

Temperature engineering unit (CF:1 in Degree C, CF:0 in Degree F)

FF:1;

Temperature engineering unit (FF:1 in Degree F, FF:0 in Degree C)

LF:0

Temperature over range flag
xxxx0xxx : In Range, Top
xxxxIxxx : Out of Range, Top
xxxxx0xx : In Range, Bottom
xxxxxIxx : Out of Range, Bottom

End

End of data string