Quick Specifications Table
Thermal Measurement Range
0 to 250°C (32 to 482°F)
Accuracy
±2% or ±2°C, whichever is greater at calibration geometry and 25°C
Repeatability
±1% or ±1°C, whichever is greater
NETD
250 mK
Thermal
Temperature Indication Resolution
0.1 (°F or °C)
Spectral Range
7-14 microns
Target Sighting
Single Laser Dot (Meets lEC Class 2 & FDA Class II requirements)
Optical Resolution
90:1
Minimum Diameter Measurement Spot
7mm (0.27”) at 60cm (24”)
Image Frame Rate
20Hz
Field of view (FOV)
17° Horizontal x 12.8° Vertical
Optical / IR
Instantaneous Field of view (IFOV)
1.9mrad
Focus
Focusable, 61 cm / 24” to infinity
Temperature Scale
°C or °F selectable
Palettes
Gray, Ironbow, or Rainbow selectable
Measurement Modes
Automatic, Semi-Automatic, or Manual selectable
Laser On/Off
9
Gain Control
9
Level Control
9
Controls
LCD Backlight
Bright, Dim, Off selectable
Adjustable Emissivity
(0.10 to 1.00 by 0.01)
9
Display Type
Liquid Crystal Display - TFT technology optimized for both indoor and
outdoor use
Reflected Background Temperature
-50 to 460°C (58 to 860°F)
Ambient Operating Temperature
0 to 50°C (32 to 122°F)
Relative Humidity
10 to 90% Non-condensing
Storage Temperature
-25 to 70°C (-13 to 158°F) [without batteries]
Storage Capacity
✑✐✐ ❉❍❁❇❅▲
Laser On Icon
9
Low Battery Icon
9
Palette Icon
9
Measurement Mode Icon
9
Thermal Analysis Software
InsideIR (included)
Operational
PC Software Operating Systems
Microsoft
®
Windows
®
98, Windows 2000 or Windows XP
Power
6 AA batteries(not included) or rechargeable battery pack (included)
Battery Life
Min. 5 hours continuous use
Data Transfer
USB interface, total transfer time up to 30s for 100 pictures
Storage Device
Flash Memory
Electrical
Tripod Mount
9
(6.35 mm (1/4) 20 unc threading)
Weight (includes batteries)
1kg (2.2 lb)
Other
Standard Accessories
• Multi-language Interactive Manual (CD ROM) • InsidelR Software •
Docking station with Universal Power Adapter and USB Connection •
Hardshell Carry Case • USB Computer Cable • Rechargeable and
Non-rechargeable Battery Packs (batteries not included) • Multi
Language Thermography Training Presentation (CD ROM) • Carrying
Pouch • Wrist Strap • Quick Reference Card
A
ccessories /
O
p
tions
Options
NIST Calibration certificate
July 1, 2003
- 8 -