15
Technical Specifications
En
gli
sh
Technical Specifications
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols Which Appear On The Device
Attention, consult accompanying documents;
Drip proof;
Type CF equipment;
Dangerous voltage;
Class II equipment;
Start/Stop;
Manufacturer.
Operating pressure range
3 to 25 cm H
2
O
Maximum single fault steady
state pressure
40 cm H
2
O
Pressure measurement
tolerance
±0.5 cm H
2
O ± 4% of the measured reading
Flow measurement tolerance
±0.1 or 20% of reading, whichever is greater
VAuto mode
4 to 25 cm H
2
O (measured at the mask); Min EPAP: 4 cm H
2
O;
Max IPAP: 25 cm H
2
O; Pressure Support: 0 to 10 cm H
2
O
S, ST and T modes
IPAP: 4 to 25 cm H
2
O (measured at the mask); EPAP: 3 to 25 cm H
2
O
(measured at the mask); Pressure Support: 0 to 22 cm H
2
O
CPAP mode
4 to 20 cm H
2
O (measured at the mask)
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
26 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2002
28 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
Sound power level
36 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007
Dimensions (L x W x H)
4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
Weight
2.9 lb (1.3 kg)
Power supply
Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)
Housing construction
Flame retardant engineering thermoplastic
Operating temperature
41
º
F to 95
º
F (+5
º
C to +35
º
C)
Operating humidity
10–95% non-condensing
Storage and transport
temperature
-4
º
F to 140
º
F (-20
º
C to +60
º
C)
Storage and transport humidity
10–95% non-condensing
Operating altitude
Sea level to 8,500’ (2,600 m)
Supplemental oxygen
Recommended maximum supplemental oxygen flow:
4 L/min (VAuto mode); 15 L/min (CPAP, S, ST and T modes)
Electromagnetic compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments
Air filter
Two-layered, powder-bonded, polyester non-woven fiber
Air tubing
Flexible plastic, 1 x 6’6” (2 m)
Air outlet
The 22 mm conical air outlet complies with ISO 5356-1
IEC 60601-1 classification
Class II (double insulation), Type CF