Retrotec Energy Innovations, Ltd.

41

DM-2A Mark II 2008-02-22

7. Press the e

nteR

keypad button and input either the Envelope Area or the Net Floor

Area as follows:

• The Envelope Area if you wish to display Air Permeability as specified in the

standard.

• The Net Floor Area is you wish to display Specific Leakage Rate as specified in

the standard.

8. Press the @ p

ReSSuRe

button to display the Flow @ 50.0 Pa, Air Changes @ 50.0 Pa, and

either Air Permeability or Specific Leakage Rate @ 50.0 Pa.

9. Results may be read directly from the DM-2A mark II.

10. In order to record both Air Permeability and Specific Leakage Rate, you will

need to re-enter the Envelope Area and Net Floor Area values according to your
measurements and/or calculations.

DM-2A Mark II Dual-Channel Pressure Gauge

User Manual

A:

Specific Configurations