Guidance and manufacturer’s declaration—electromagnetic immunity

The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or

the user of the S8 device should assure that the device is used in such an environment.

Immunity test

IEC60601-1-2
test level

Compliance
level

Electromagnetic environment—guidance

Portable and mobile RF communications

equipment should be used no closer to any part

of the S8 device, including cables, than the

recommended separation distance calculated

from the equation applicable to the frequency of

the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

3 Vrms

d = 1.17

P

Radiated RF

IEC 61000-4-3

10 V/m

80 MHz to 2.5 GHz

10 V/m

d = 0.35

P 80 MHz to 800 MHz

d = 0.70

P 800 MHz to 2.5 GHz

where P is the maximum output power rating of

the transmitter in watts (W) according to the

transmitter manufacturer and d is the

recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in each

frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site

survey should be considered. If the measured field strength in the location in which the S8 device is used exceeds

the applicable RF compliance level above, the S8 device should be observed to verify normal operation. If

abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the

S8 device.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.