Leaning, Aintenance – ResMed S8 User Manual
Page 26

21
E
NG
LISH
C
LEANING
AND
M
AINTENANCE
C
LEANING
AND
M
AINTENANCE
You should regularly carry out the cleaning and maintenance described in this section.
Refer to your mask manual for detailed instructions.
D
AILY
C
LEANING
Mask
Clean the mask according to the instructions supplied with the mask.
Air tubing
Disconnect the air tubing from the
S8 E
SCAPE
(and humidifier, if used) and hang the
tubing and mask in a clean, dry place until next use.
Humidifier
If you are using a humidifier, clean it according to the instructions in the humidifier
user’s manual.
W
EEKLY
C
LEANING
1. Remove the air tubing from the
S8 E
SCAPE
and the mask.
2. Wash the air tubing in warm water using mild detergent. Rinse thoroughly, hang
and allow to dry.
3. Before next use, assemble the mask and headgear according to the user
instructions.
4. Reconnect the air tubing to the air outlet and mask.
P
ERIODIC
C
LEANING
1. Clean the exterior of the
S8 E
SCAPE
with a damp cloth and mild detergent.
2. Check the air filter for blockage or holes (see “Replacing the Air Filter” on
page 22).
!
WARNING
•
Beware of electric shock. Do not immerse the unit or power cord in water.
Always unplug the unit before cleaning and be sure it is dry before plugging back
in.
•
The mask system and air tubing are subject to normal wear and tear. Inspect
them regularly for damage.